Figure 1.

Figure 1 shows a comparison of the present "multi-station" technique
with the new continuously moving table technique developed at the MR
Research Lab.In the case of the former, three static stations are imaged
separately and the table is moved one FOV between scans.This technique
can be inefficient because significant time is required to move to the
next FOV and set up steady state conditions.On the right, the continuously
moving table technique acquires the same total FOV in one smooth motion,
nominally moving from torso to feet along with the contrast bolus.
|